Q What is the lowest flow rate that can be achieved with the micropump mp6?

The lowest flow rate is very dependent on the conditions in the customers test system. Generally a flow rate of about 20 µl/min for liquids can be achieved in most cases. With thorough optimization much lower rates can be achieved. In order to ensure stable pump performance, if possible the frequency should be lowered first and the amplitude should be kept as high as possible. If necessary a precision orifice can be used to limit the outlet flow additionally.